Twister Spray Chamber with Helix CT

Glass Expansion SKU: 20-809-0222HE

Call for price

Available Now!

Twister Spray Chamber with Helix CT  - for semi-demountable torch, 50ml cyclonic, Borosilicate glass 

Glass Expansion Part Number: 20-809-0222HE
Equivalent to OEM Part Numbers: 33010015, 1147929063, 13-410-414, G8000-70007

Includes: 

  • UFT-16-75 and UFT-2-700 UniFit Connectors (1 each)
  • 70-803-1439 Helix CT locking screw with seal (1 unit)

Suitable for the Following Instruments:

  • Agilent (Varian) | MP-AES | 4100 / 4200 MP-AES | Radial
  • Jobin Yvon (Horiba) | ICP-OES | 24, 38 & 138 Series, Ultima, Activa | Radial
  • Leeman (Teledyne) | ICP-OES | Leeman Series Radial, Prodigy, Profile | Radial
  • Nu Instruments | ICP-MS | Nu Plasma
  • PerkinElmer | ICP-OES | Optima 3000RL, 3100RL, 3200RL | Radial
  • Shimadzu | ICP-OES | 7000, 8000, 9000, 9800 Series | Radial / Axial
  • Spectro (Ametek) | ICP-OES | SOP (Modula, CIROS, Genesis, ARCOS) | Radial
  • Thermo | ICP-MS | PlasmaQuad, Axiom
  • Thermo | ICP-OES | Iris, Atomscan 16 & 25, ICAP & 61E, Intrepid | Radial

Twister cyclonic - reduces solvent load

The Twister cyclonic spray chamber features a central transfer tube which acts as a secondary particle separator helping separate larger aerosol particles from the sample. This reduces solvent load in the plasma without compromising detection limits.

General Specifications:

  • Material: Borosilicate glass
  • Volume: 50ml
  • Physical reproducibility ~1%
  • Low RSD's - highly accurate construction similar to Tracey cyclonic, except with central transfer tube
  • Application uptake range 0.2 to 3.0ml/min
  • Produces smaller mean droplet size

Jacketed Twister spray chamber (borosilicate glass material)

The Jacketed Twister spray chamber is made with an external jacket for the introduction of coolant for temperature control. Thermostatting to constant temperature helps ensure highly reproducible results and long term stability. Can also be cooled to suppress volatilization, allowing the direct aspiration of highly volatile solvents such as naphtha and petrol/gasoline. Recommended for analyses of organics samples.